Product Description

Supreme

Reshaping the sterilization world. With its unrivalled low consumption, Supreme is the new market benchmark.

Water Treatment

The ceramic recycle filter, incorporated into the secondary tank, makes it possible to distil and purify water so that it can be reused in the following sterilization cycle. The built-in conductivity sensor ensures compliance with water quality parameters as it protects the machine internal components from excess lime. Total consumption is 5 litres per 50 cycles, meaning approx. 100 ml/cycle, a real revolution.

Operating Principle

Unlike traditional sterilizers, Supreme does not require any external water connection for mains water treatment. Thanks to the demineralisation filter on the main tank, the user can fill the latter with plain tap water, thereby significantly optimising spaces, streamlining processes and achieving considerable savings. Supreme’s different filtration systems guarantee unprecedented ease of operation.

Supreme: 50 sterilization cycles with just 5 litres of tap water. Consumption is approx. 100 ml/cycle.

Di.V.A. - Digital Virtual Assistant

Di.V.A. stands for Digital Virtual Assistant. It is the first cloud-based control and traceability system for MOCOM sterilization devices. Features include access to tutorial videos, display of executed cycles, check on machine efficiency and status, reading of machine usage statistics, and monitoring the frequency of sterilization tests.

Remote Technical Support with Easy Check

All sterilizers of the Supreme range are equipped with incorporated WiFi and Ethernet port. Connected to the Internet by enabling the Easy Check service, they can receive remote support, significantly reducing intervention times.

Printers and Traceability Software

The Supreme range can be completed with 2 printers available as optional accessories and MyTrace, Cefla’s traceability software supplied with Supreme, which completes the sterilization process by associating each set of sterilized instruments to the patient through a bar code.

NFC - Near Field Communication

Improves workflow by allowing assistants to register with a special user card instead of a PIN code, providing access to machine functions for authorized personnel only.

Light Communication and Video Tutorial

Communicates through colours with a door built-in LED bar to provide immediate information about the machine status. Also offers video tutorials and important information for using and maintaining the machine through its colour display.

Specifications

Accessories

  • External printer
  • Network printer
  • Front filling kit
  • Automatic filling kit
  • SV aux kit
  • Modular tray holder
  • Bar code reader

Technical Data

Supreme 17 Supreme 22 Supreme 28
Power supply voltage 220/240V 50Hz 220/230V 60Hz 220/240V 50Hz
Rated power 2300 W
External dimensions (W x H x D) 480 x 500 x 600 mm
Chamber dimensions (Diam. x Depth) 250 x 350 mm 250 x 450 mm 280 x 450 mm
Total weight 54 kg 55 kg 56 kg
Tank capacity 4.5 litres
Autonomy (with max. water level and both filters installed) From 40 to 50 cycles From 40 to 50 cycles From 40 to 50 cycles

Sterilization Cycles

Cycle Type Sterilization Time Supreme 17 Supreme 22 Supreme 28 Drying *
134°C UNIVERSAL B 4 min 27 min 29 min 37 min 13-17 min
121°C UNIVERSAL B 20 min 43 min 46 min 50 min 13-17 min
134°C VELOCE S 4 min 20 min 23 min 26 min 1 min
134°C WRAPPED SOLID INSTRUMENTS S 4 min 18 min 22 min 26 min 13-17 min
134°C PRION B 18 min 41 min 43 min 51 min 13-17 min
XXX°C CUSTOM S Customizable Customizable Adjustable
VACUUM TEST TEST 18 min 18 min 19 min
HELIX / B&D TEST TEST 20 min 24 min 28 min
VACUUM + HELIX / B&D TEST TEST 42 min 46 min 51 min

* Drying varies according to machine model and volume. Note: times do not take the max. pre-heating time into account (10 min.). Note: times may vary depending on the load and power supply. Note: 120 V model times may undergo increments up to a max. of 20%.